HomeCleanroomMicrobiology Room EquipmentBiosafety Cabinet
BSC Class II · EN 12469 · NSF 49 · Type A2 · Three-Way Protection

Biosafety Cabinet (BSC Class II)

Simultaneous Protection for Personnel, Product & Environment — In One Ventilated Enclosure

The Metrolabs Class II Biosafety Cabinet is the structural cornerstone of any microbiology room that handles potentially infectious biological agents. Using high-velocity laminar downward airflow and dual HEPA filtration, it creates a ventilated work zone that simultaneously shields the operator from bio-aerosol exposure, protects samples from cross-contamination, and purifies exhaust air before returning it to the laboratory — a three-way containment performance that no other single piece of laboratory equipment provides.

Class II Type A2 HEPA 99.997% — Supply & Exhaust EN 12469 / NSF 49 SS 304 / SS 316L Coved Interior UV-C Germicidal Microprocessor Controller
Metrolabs Biosafety Cabinet Class II – HEPA filtration, three-way protection, SS coved interior, pharmaceutical microbiology
🧪 BSC Class II
Three-way protection · HEPA H14 · UV-C · Microprocessor controller
Metrolabs BSC – sloped sash, recessed grille, airflow visualization
🌬 Airflow
Sloped sash · Recessed grille
Metrolabs Biosafety Cabinet Class II installed in microbiology laboratory
🏭 Installed
Zero-Ledge · Flush-mounted
👤
01
Personnel Protection

Inward air curtain at ≥0.38 m/s draws room air into the cabinet through the front sash opening — preventing bio-aerosols generated at the work surface from reaching the operator’s face and breathing zone.

🧪
02
Product Protection

HEPA-filtered vertical downflow air (Grade A — ISO Class 5) sweeps the work surface with ultra-clean air — preventing ambient airborne micro-organisms from contaminating cell cultures, media, and sensitive samples.

🏭
03
Environment Protection

All exhaust air passes through a HEPA filter before returning to the laboratory — ensuring that biological agents from the work zone never escape to the surrounding room, protecting colleagues and the facility.

🔌
HEPA 99.997% — Both Supply and Exhaust
🌬
Inflow ≥0.38 m/s — EN 12469 / NSF 49
🔧
SS 304 / 316L Coved Interior — Zero-Joint
UV-C Germicidal — Sash-Interlocked
💻
Microprocessor — Velocity Alarm + ΔP
🔥
VHP Compatible — All Components Rated
BSC Class II Type A2 — Airflow Principle Diagram
⬆ EXHAUST HEPA FILTER
99.997% — purified exhaust returned to lab room (~30%)
SUPPLY HEPA FILTER + MOTOR BLOWER
H14 99.997% — downflow Grade A · 0.25–0.38 m/s
☢ UV-C GERMICIDAL LAMP
Idle period only — sash-interlocked auto-cycle
▼ HEPA downflow · Grade A / ISO Class 5
WORK ZONE — SS 304/316L Coved Surface
ISO Class 5 / Grade A clean work surface
▼ into recessed front grille ◄ room air ≥0.38 m/s
FRONT SASH — 10° Sloped Safety Glass
Inward air curtain · operator’s arms enter here
CONTAMINATED AIR SPLITS AT INTAKE GRILLE
~70% recirculates → supply HEPA → work zone · ~30% exhausted → exhaust HEPA → lab room
💻 MICROPROCESSOR · 🦵 HEIGHT-ADJUSTABLE SS STAND (opt.)
Velocity + ΔP + UV hours monitoring
🌬 Inflow Velocity
≥0.38 m/s (75 fpm) — NSF 49 minimum for personnel protection.
⬇️ Downflow Velocity
0.25–0.38 m/s — HEPA Grade A ISO Class 5 across work surface.
🔌 HEPA Split
~70% recirculates · ~30% purified through exhaust HEPA.
🔋 Vibration
Suspension-isolated blower — stable for balance & microscopy.
Step-by-Step Operation

How the Metrolabs BSC Class II Protects All Three Parties Simultaneously

The BSC Class II is the only laboratory equipment that simultaneously protects the operator, the sample, and the laboratory environment. Understanding how it achieves all three is essential for correct use and for selecting the right unit for your application.

1

Room Air Drawn In — Personnel Protection Begins

When the cabinet is operating, a fan mounted at the top of the unit draws air inward through the front aperture (the sash opening where the operator works). This inward airflow — maintained at a minimum of 0.38 m/s (75 fpm) per NSF 49 — creates a persistent air curtain at the sash opening that continuously sweeps any bio-aerosols generated at the work surface away from the operator’s face and breathing zone and into the cabinet interior. As long as this inward air velocity is maintained above the threshold, no bio-aerosol from the work zone can travel outward through the sash opening to the operator.

2

HEPA Downflow Creates Grade A Work Zone — Product Protection

Simultaneously, the fan draws air from the cabinet plenum through the supply HEPA filter (H14, 99.997% at 0.3 micron MPPS) and delivers it downward across the work surface at 0.25–0.38 m/s in a vertical laminar pattern. This constant supply of HEPA-filtered air maintains the work surface environment at Grade A (ISO Class 5) — fewer than 3,520 particles per cubic metre at 0.5 micron — preventing ambient microorganisms from contaminating cell cultures, media, and samples on the work surface.

3

Contaminated Air Splits at Front Grille — 70% / 30% Split

The downflow air — which has now swept across the work surface and may carry biological agents from the work in progress — mixes with the inflow air at the front intake grille and is drawn into the recirculation and exhaust plenum below the work surface. Here the contaminated air is split: approximately 70% is directed back to the supply HEPA filter for recirculation, and approximately 30% is directed to the exhaust HEPA filter for purification before discharge back to the laboratory room. This Type A2 split — 70/30 recirculation/exhaust — is the standard for non-volatile chemical applications.

4

Exhaust HEPA Purifies Room Air — Environment Protection

The exhaust portion of the contaminated air passes through the exhaust HEPA filter (H14, 99.997% at 0.3 micron) before being returned to the laboratory room. This exhaust HEPA removes 99.997% of all biological agents — bacteria, viral particles, fungal spores, and aerosolised cells — ensuring that no living biological agent escapes from the BSC interior to the surrounding laboratory environment. The laboratory personnel working outside the BSC are protected from biological agents generated inside it throughout the working day.

5

Microprocessor Monitors Safety — Alarm Before Operator Awareness

The microprocessor controller continuously samples both inflow and downflow air velocity sensors at a high frequency — detecting a drop in inflow velocity (which would compromise personnel protection) before the velocity has dropped far enough for the operator to notice a change in the airflow sensation at the sash opening. When inflow velocity falls below the safe threshold (0.38 m/s), an audible alarm and visual alert activate immediately — alerting the operator to stop work, close the sash, and investigate the cause before any biological material has escaped the cabinet.

Classification Guide

BSC Classes & Types — Choosing the Right Level of Containment

The type of BSC required depends on the biosafety level of the organisms being handled, whether chemical vapours are present, and whether product protection is needed:

Class I

Personnel & Environment — No Product Protection

Inward airflow only. Protects the operator and the room from bio-aerosols. No HEPA supply — unfiltered room air flows over the work surface. Suitable for general microbiology where sample sterility is not required. Rarely used in modern pharmaceutical laboratories.

Class II Type A2 — Most Common — Metrolabs Standard

Personnel + Product + Environment — BSL-1 and BSL-2

Recirculates ~70% of air through supply HEPA, exhausts ~30% through exhaust HEPA. Inflow ≥0.38 m/s. Grade A work surface. Used in ~90% of all BSC installations worldwide. Suitable for cell culture, microbiology, pharmaceutical QC, hospital labs. Standard for BSL-1 and BSL-2. EN 12469 / NSF 49 certified.

Class II Type B2 — Volatile Chemicals

100% Total Exhaust — Chemical + Biological Dual Hazard

100% of cabinet air is exhausted through a dedicated external duct — no recirculation. Required when volatile toxic chemicals, radionuclides, or carcinogens are used alongside biological work inside the cabinet. Requires external ducting installation — not a plug-and-play cabinet.

Class III

Gas-Tight Glovebox — BSL-3/4 Maximum Containment

Completely sealed, gas-tight enclosure under negative pressure. Operator works through heavy-duty rubber gloves attached to ports. All supply air through two HEPA filters in series. Exhaust through two HEPA filters in series. For BSL-3 and BSL-4 — highest-risk pathogens (Ebola, SARS, anthrax).

Metrolabs Biosafety Cabinet Class II – ergonomic sloped sash, large work zone, microprocessor display, LED illumination
Design Detail

Ergonomic Design — Built for Extended Biological Work Sessions

Biosafety cabinets are used for hours on end. Metrolabs BSC Class II units incorporate ergonomic design features that reduce operator fatigue, improve visibility, and maintain the protective air curtain throughout extended work sessions:

⬇️
10° Sloped Front SashEnables a better seated position and wider field of view into the work zone without requiring the operator to lean forward — maintaining distance from the protective air curtain.
🔧
Recessed Air Intake GrilleThe front air intake grille is recessed below the sash opening — preventing the operator’s forearms from blocking the inward air curtain when working at the front of the work surface.
💡
LED Task LightingBright, low-heat LED lighting illuminates the entire work surface uniformly — no shadows, no heat emission that affects temperature-sensitive cultures, and no UV component that would compromise personnel.
💻
Forearm Support / ArmrestAn ergonomic forearm support above the intake grille allows the operator to rest their forearms during extended pipetting tasks without fatigue and without placing weight on the grille that would obstruct airflow.
Standard Features

Six Engineering Features of the Metrolabs BSC Class II

Every Metrolabs Biosafety Cabinet Class II is built to these standards — making it a precision primary containment instrument, not simply a ventilated enclosure:

🔌

Dual H14 HEPA Filtration — 99.997% — Supply + Exhaust

Two H14 HEPA filters (EN 1822 rated — 99.997% minimum efficiency at 0.3 micron MPPS) are installed in the BSC: one on the supply air path and one on the exhaust air path. Both are installed in gel-seal or knife-edge frames that prevent any unfiltered air from bypassing the filter perimeter. A factory DOP/PAO integrity test is performed on both HEPA filters before the unit is shipped, and the test report is supplied as part of the IQ documentation. Differential pressure gauges across both filters provide visible filter loading indicators. A G4 pre-filter in the intake path extends HEPA service life.

H14 SupplyH14 Exhaust99.997%Gel-Seal FrameDOP Factory Test
🔧

Coved SS Interior — Zero-Joint, Single-Piece Construction

The internal work zone is fabricated from surgical-grade SS 304 (standard) or SS 316L (optional) with a single-piece coved interior — all internal corners and floor-to-wall junctions radiused, with no internal weld joints or seam lines. This eliminates the standard point of microbial colonisation (the weld bead at the floor-to-wall junction), leaving a continuous smooth surface at every corner. The electropolished or mirror-polished internal surface (Ra <0.5μm) makes contamination immediately visible under the LED work zone lighting — supporting visual inspection-based cleaning verification.

SS 304 / 316LSingle-Piece CovedEP Ra <0.5μmVHP Compatible
💻

Microprocessor Controller — Real-Time Safety Monitoring with Velocity Alarm

The microprocessor-based controller samples inflow and downflow air velocity sensors continuously — at a frequency that detects a transient airflow anomaly within seconds. The inflow velocity alarm activates when inflow falls below the NSF 49 minimum threshold (0.38 m/s / 75 fpm), producing an audible and visual warning before the operator is aware of any change. The display provides real-time readings of inflow velocity, downflow velocity, HEPA filter differential pressure, UV-C lamp cumulative hours, and system status. A USB data port allows logging for GMP environmental monitoring and audit records.

Velocity AlarmΔP DisplayUV HoursUSB Log Port

UV-C Germicidal Lamp — Safety-Interlocked Automated Decontamination

The integrated UV-C germicidal lamp provides full coverage of the work surface and side walls during idle periods between work sessions. The lamp cannot be activated when the sash is in the open operating position — the sash-position interlock ensures personnel are never exposed to UV-C radiation. If the sash is opened during an active cycle, the lamp deactivates immediately. The controller tracks cumulative UV-C hours and alerts when the lamp approaches end-of-life (typically 7,000–9,000 hours). It supplements, but does not replace, the manual IPA wipe-down required at the start and end of every session.

Sash-InterlockedAuto-DeactivateHour CounterProgrammed Timer
🔥

VHP Compatibility — All Components Tested for Room-Wide Decontamination

All internal components — silicone sash sealing strips, front viewing panel glazing, UV-C lamp end caps, HEPA filter housing gaskets, and work surface drain valve seals — are specified and tested to withstand room-wide Vaporised Hydrogen Peroxide sterilisation cycles at 1000 ppm for standard pharmaceutical durations. For laboratories using room-wide VHP decontamination, this ensures the BSC can be processed in the same cycle as the room without equipment damage — returning to service immediately after aeration without BSC-specific post-VHP inspection.

1000 ppm VHPAll GasketsViewing PanelDrain Seals
🔋

Vibration-Isolated Blower — Stable Work Surface for Precision Biology

The motor-blower is mounted with vibration-isolation suspension — calibrated coil springs and damping washers that decouple blower vibration from the cabinet body and work surface. This is essential for two classes of biological work: analytical balance weighing of culture media components (where surface vibration above 0.01g affects repeatability) and high-magnification optical microscopy (where vibration above ~0.5 micron amplitude causes image blur above 200x). The vibration-isolated blower also reduces acoustic noise at the operator position, contributing to a quiet environment for extended concentrated work.

Spring IsolationBalance StableLow AcousticMicroscopy OK
Technical Specifications

BSC Class II — Standard Technical Specification

Metrolabs BSC Class II standard parameters — customisable at every dimension and feature:

SpecificationDetails
ClassificationClass II Type A2
Compliance Standard✓ CertifiedEN 12469 / NSF/ANSI 49
Inflow Velocity≥ 0.38 m/s (75 fpm) — NSF 49 minimum
Downflow Velocity0.25 – 0.38 m/s — Grade A ISO Class 5
Air Recirculation Split~70% recirculated / ~30% exhausted
Supply HEPAH14 — 99.997% at 0.3μm — EN 1822
Exhaust HEPAH14 — 99.997% at 0.3μm — EN 1822
Interior MOCSS 304 std / SS 316L optional — coved
Front Sash10° sloped safety glass — recessed grille
UV-C LampStandard — sash-interlocked — programmable timer
ControllerMicroprocessor — velocity alarm — USB log
Standard Widths900 / 1200 / 1500 / 1800 mm
Power Supply220 V / 50 Hz
Full Specification
ClassClass II Type A2
StandardEN 12469 / NSF/ANSI 49
Inflow Min≥0.38 m/s (75 fpm)
Downflow0.25–0.38 m/s Grade A
Recirculation~70% supply / ~30% exhaust
HEPA FiltersH14 — 99.997% — supply + exhaust
G4 Pre-FilterIntake path — extends HEPA life
Interior MOCSS 304 std / SS 316L opt
Interior FinishEP Ra <0.5μm / Mirror
Sash10° sloped — recessed grille
UV-CStandard — sash-interlocked
ControllerMicroprocessor + USB log
Standards & Compliance
EN 12469NSF/ANSI 49Class II A2EN 1822 HEPAWHO LBMASTM A967GMP / ISO
Optional Configurations

Optional Upgrades & Configurations

🧱

SS 316L Interior Upgrade

Upgrade from SS 304 to SS 316L for superior corrosion resistance against VHP and chloride disinfectants in Grade A/B pharmaceutical BSC applications.

📈

Height-Adjustable Stand

Motorised or manual height-adjustable stand for sitting or standing operation — with integrated ergonomic footrest — for operator comfort during extended sessions.

🔍

PAO / DOP Test Ports

Dedicated upstream and downstream sampling ports for PAO filter integrity test during commissioning and routine ISO/GMP annual certification — required for qualification documentation.

💾

USB / BMS Integration

USB port for operational parameter data download — or Ethernet/BMS integration to relay airflow velocity, filter pressure, and UV hours to the facility building management system.

🕐

UV-C Hour Meter

Cumulative UV-C lamp hour counter on microprocessor display for planned replacement scheduling — ensuring germicidal output remains above effective threshold throughout service life.

Exhaust Canopy Connection

Thimble connection canopy allowing the BSC to be connected to the building exhaust system for work involving small quantities of volatile chemicals alongside BSL-1/2 biological agents.

📋

Calibration Certificate

Traceable calibration certificate for the airflow velocity sensor and display — required for GMP instrument qualification documentation at installation and annual recertification.

🚨

Sash Open Alarm

Audible alarm activated when sash is raised above the safe working height — alerting the operator that the protective air curtain may be compromised before biological work commences.

Applications

BSC Class II Applications Across Every Regulated Biological Laboratory

EN 12469 / NSF 49 Class II A2 Biosafety Cabinets for every environment where three-way protection — personnel, sample, and environment — is required simultaneously:

💊

Pharmaceutical QC Microbiology

BSC Class II for sterility testing, microbial limit testing, environmental monitoring plate incubation, and cell culture work. GMP IQ documentation at handover.

🧪

Cell Culture & Biotech

BSC Class II for mammalian cell culture, hybridoma work, viral vector production, and primary cell isolation — protecting both the operator and the culture simultaneously.

BSL-2 Microbiology

BSC Class II for work with moderate-hazard agents — Staphylococcus, Salmonella, hepatitis B, HIV — at Biosafety Level 2. WHO LBM and CDC/NIH BMBL compliant.

🏥

Hospital & Clinical Microbiology

BSC Class II for clinical sample processing — blood cultures, sputum specimens, MRSA/VRE cultures, and infectious sample handling. NABH compliant throughout.

🍽️

Food Microbiology Testing

BSC Class II for FSSAI food microbiology — pathogen testing (Salmonella, Listeria, E. coli O157), sterility confirmation, and microbiological quality testing.

🔬

DNA & PCR Work

BSC Class II for DNA recombinant work, gene therapy research, and PCR preparation — the downflow clean air prevents amplified DNA from contaminating subsequent reactions.

💊

Pharmacy Drug Compounding

BSC Class II for aseptic pharmaceutical compounding — providing the sterile Grade A environment required while protecting compounding personnel from biological contamination.

🔬

Research & NABL Labs

BSC Class II for NABL-accredited biotechnology and microbiology research laboratories. Annual NSF-certified airflow recertification. VHP compatible. Full NABL documentation.

The Metrolabs Advantage

High-Accuracy Fabrication for Long-Term Biosafety Performance

Metrolabs focuses on the high-accuracy fabrication, advanced control systems, and comprehensive documentation that ensures your BSC meets global biosafety standards throughout its service life:

🔌

Factory DOP Test on Both HEPA Filters Before Dispatch

Every Metrolabs BSC undergoes a factory DOP/PAO integrity test on both the supply and exhaust HEPA filters — confirming zero bypass at the filter perimeter before the unit leaves the factory. The test report is included in the IQ documentation package. A second DOP test at site confirms integrity is maintained through shipping and installation.

🌬

Certified Airflow Velocity — Inflow + Downflow Confirmed at Handover

Airflow velocity measurements at multiple work zone positions are performed and documented at handover — confirming that both the personnel protection threshold (≥0.38 m/s inflow) and the Grade A downflow velocity are achieved and distributed uniformly before the unit enters service.

🔥

VHP Compatible Throughout — Every Component, Every Cycle

All components — gaskets, sash seal, UV-C end caps, HEPA housing seals, drain valve — are tested for VHP compatibility at 1000 ppm. Laboratories using VHP room decontamination can process the BSC in the same cycle as the room, returning to service after aeration without disassembly.

📝

Annual Recertification Support — NSF Annual Inspection Requirement

BSCs must be annually recertified by a trained third-party NSF/EN certifier. Metrolabs provides the DOP test ports (optional) and calibration certificates required by the recertification procedure, and can coordinate with accredited recertification services for annual inspection scheduling.

💡 Technical Insight — Annual Airflow Visualisation (Smoke Test)
To maintain the highest level of biosafety, Metrolabs recommends performing an Airflow Visualisation (Smoke Test) as part of the annual NSF recertification. A smoke tracer introduced at the sash opening confirms that the inward air curtain maintains a clean sweep across the full width of the sash without any turbulent eddy that could carry bio-aerosol outward. Smoke at the perimeter of the downflow zone reveals any disruption to the laminar pattern caused by items on the work surface — a common finding when tall flasks or stacked equipment create turbulent wakes. The annual smoke test catches the airflow patterns that the velocity sensors alone cannot confirm — because velocity is measured at fixed points, not across the full three-dimensional work zone geometry.

BSC Class II vs Laminar Air Flow (LAF) — Key Differences

Feature 🧪 BSC Class II 🔌 LAF Unit
Protects operator
✓ Inward air curtain
✕ No protection
Protects sample
✓ HEPA downflow
✓ HEPA unidirectional
Protects environment
✓ HEPA exhaust
✕ Air flows to room
Infectious agents
✓ BSL-1 and BSL-2
✕ Never — hazardous
Compliance standard
EN 12469 / NSF 49
ISO 14644-1 Class 5
Typical applications
Pathogens, cell cultures, BSL work
Media prep, PCR, electronics

A Laminar Air Flow unit must never be used for work with infectious agents, pathogens, or biological material with aerosol hazard potential — it does not protect the operator. Use a BSC Class II whenever the material being handled has any potential to cause infection or allergic sensitisation through the respiratory route. See LAF Units →

Supply & Commissioning

From Specification to Factory-Tested & IQ-Documented BSC

Metrolabs manages the complete BSC supply and commissioning process:

01
📋

Specification & Biosafety Assessment

Biosafety level confirmed (BSL-1 / BSL-2). Agents handled reviewed. MOC (SS 304 / SS 316L), cabinet width (900–1800mm), optional features, and room ventilation requirements confirmed. EN 12469 / NSF 49 compliance confirmed. Dimensioned drawing produced and signed off.

02
🔌

Precision Fabrication

Cabinet body fabricated — SS 304 or SS 316L coved single-piece interior. HEPA plenum and blower installed with vibration mounts. Supply and exhaust H14 HEPA filters installed in gel-seal frames. G4 pre-filter fitted. Microprocessor controller wired. UV-C lamp mounted with sash interlock. LED lighting fitted. Passivated + polished.

03
🌬

Factory Test & Validation

Airflow velocity measured — inflow ≥0.38 m/s · downflow 0.25–0.38 m/s · DOP/PAO integrity test on both HEPA filters · UV-C sash interlock tested · Microprocessor alarm tested at threshold · USB data log verified · Smoke visualisation test. All results documented and test report issued with unit.

04

IQ Documentation Package

SS MTCs, ASTM A967 passivation cert, H14 EN 1822 filter cert, DOP/PAO test report (supply + exhaust HEPA), velocity measurement grid, alarm test record, UV-C interlock test record, smoke test record, calibration certificates, and installation photographs compiled for GMP/ISO laboratory IQ documentation at handover.

Visit Us
Metrolabs — Chennai, Tamil Nadu
📍

Factory & Office

4, SIDCO-MICRO, Thirumudivakkam, Chennai – 600 132, Tamil Nadu

🗺 Open in Google Maps ↗
🌎

Website & GSTN

www.metrolabs.biz www.schoollabfurniture.com

GSTN: 33AASFM3382K1ZT

Get a Free Quote

Ready to Install a Biosafety Cabinet Class II in Your Laboratory?

Contact Metrolabs for a free biosafety assessment. Our specialists will confirm your BSL level, agent type, cabinet width, optional features, and documentation requirements — then fabricate, factory-validate, install, and commission your BSC with complete IQ documentation at delivery.

🔌
Dual H14 HEPA — 99.997% Supply + Exhaust — Factory DOP Test
Grade A downflow · HEPA exhaust · EN 1822 certification
🌬
EN 12469 / NSF 49 — Inflow ≥0.38 m/s — Velocity Alarm
Microprocessor continuous monitoring · Alarm before awareness
🔧
Coved SS 304/316L Interior — Zero-Joint, Zero-Ledge
Single-piece pressed · EP Ra <0.5μm · VHP 1000 ppm
📋
Audit-Ready — DOP Test + Velocity Grid + IQ Documentation
Both HEPA filters tested · GMP/ISO qualification ready

🧪 Request a Free BSC Class II Quote