Biosafety Cabinet (BSC Class II) – Three-Way Personnel, Product & Environment Protection | Metrolabs Chennai
Home Cleanroom Microbiology Room Equipment Biosafety Cabinet
BSC Class II · EN 12469 · NSF 49 · Type A2 · Three-Way Protection

Biosafety Cabinet
(BSC Class II)

Simultaneous Protection for Personnel, Product & Environment — In One Ventilated Enclosure

The Metrolabs Class II Biosafety Cabinet is the structural cornerstone of any microbiology room that handles potentially infectious biological agents. Using high-velocity laminar downward airflow and dual HEPA filtration, it creates a ventilated work zone that simultaneously shields the operator from bio-aerosol exposure, protects samples from cross-contamination, and purifies exhaust air before returning it to the laboratory — a three-way containment performance that no other single piece of laboratory equipment provides.

Class II Type A2 HEPA 99.997% — Supply & Exhaust EN 12469 / NSF 49 SS 304 / SS 316L Coved Interior UV-C Germicidal Microprocessor Controller
Metrolabs Biosafety Cabinet Class II – HEPA filtration, three-way protection, SS coved interior, pharmaceutical microbiology
🧪 BSC Class II

Three-way protection · HEPA H14 · UV-C · Microprocessor controller
Metrolabs BSC – sloped sash, recessed grille, airflow visualization
🌬 Airflow
Sloped sash · Recessed grille
Metrolabs Biosafety Cabinet Class II installed in microbiology laboratory
🏭 Installed
Zero-Ledge · Flush-mounted
👤
01
Personnel Protection

Inward air curtain at ≥0.38 m/s draws room air into the cabinet through the front sash opening — preventing bio-aerosols generated at the work surface from reaching the operator’s face and breathing zone.

🧪
02
Product Protection

HEPA-filtered vertical downflow air (Grade A — ISO Class 5) sweeps the work surface with ultra-clean air — preventing ambient airborne micro-organisms from contaminating cell cultures, media, and sensitive samples.

🏭
03
Environment Protection

All exhaust air passes through a HEPA filter before returning to the laboratory — ensuring that biological agents from the work zone never escape to the surrounding room, protecting colleagues and the facility.

🔌
HEPA 99.997% — Both Supply and Exhaust
🌬
Inflow ≥0.38 m/s — EN 12469 / NSF 49
🔧
SS 304 / 316L Coved Interior — Zero-Joint
UV-C Germicidal — Sash-Interlocked Auto-Cycle
💻
Microprocessor — Velocity Alarm + Filter ΔP
🔥
VHP Compatible — All Components Rated
BSC Class II Type A2 — Airflow Principle Diagram
EXHAUST HEPA FILTER 99.997% — purified exhaust to lab room SUPPLY HEPA FILTER + MOTOR BLOWER H14 99.997% — downflow Grade A — 0.25–0.38 m/s UV-C GERMICIDAL LAMP (idle period — sash-interlocked) HEPA downflow ↪ Grade A / ISO Class 5 SS 304/316L COVED WORK SURFACE (Grade A) WORK ZONE ISO Class 5 / Grade A FRONT INTAKE GRILLE — Recessed (operator arm does not block air curtain) ROOM AIR inflow ≥0.38 m/s FRONT SASH 10° sloped safety glass — inward air curtain Operator’s arms enter here CONTAMINATED AIR SPLITS AT INTAKE GRILLE ~70% recirculates → supply HEPA → work zone again ~30% exhausted → exhaust HEPA → laboratory room MICROPROCESSOR Velocity + ΔP + UV hours HEIGHT-ADJUSTABLE SS STAND (optional) Personnel: inward curtain ≥0.38 m/s · Product: HEPA downflow 0.25–0.38 m/s · Environment: HEPA exhaust ~70% recirculation · ~30% exhausted · Type A2 — EN 12469 / NSF 49
🌬 Inflow Velocity
≥0.38 m/s (75 fpm) — NSF 49 minimum safe threshold for personnel protection.
⬇️ Downflow Velocity
0.25–0.38 m/s — HEPA-filtered Grade A ISO Class 5 air across work surface.
🔌 HEPA Split
~70% recirculates through supply HEPA · ~30% purified through exhaust HEPA.
🔋 Vibration
Suspension-isolated blower — work surface stable for balance, microscopy.
Step-by-Step Operation

How the Metrolabs BSC Class II Protects All Three Parties Simultaneously

The BSC Class II is the only laboratory equipment that simultaneously protects the operator, the sample, and the laboratory environment. Understanding how it achieves all three is essential for correct use and for selecting the right unit for your application.

1

Room Air Drawn In — Personnel Protection Begins

When the cabinet is operating, a fan mounted at the top of the unit draws air inward through the front aperture (the sash opening where the operator works). This inward airflow — maintained at a minimum of 0.38 m/s (75 fpm) per NSF 49 — creates a persistent air curtain at the sash opening that continuously sweeps any bio-aerosols generated at the work surface away from the operator’s face and breathing zone and into the cabinet interior. As long as this inward air velocity is maintained above the threshold, no bio-aerosol from the work zone can travel outward through the sash opening to the operator.

2

HEPA Downflow Creates Grade A Work Zone — Product Protection

Simultaneously, the fan draws air from the cabinet plenum through the supply HEPA filter (H14, 99.997% at 0.3 micron MPPS) and delivers it downward across the work surface at 0.25–0.38 m/s in a vertical laminar pattern. This constant supply of HEPA-filtered air maintains the work surface environment at Grade A (ISO Class 5) — fewer than 3,520 particles per cubic metre at 0.5 micron — preventing ambient microorganisms from contaminating cell cultures, media, and samples on the work surface.

3

Contaminated Air Splits at Front Grille — 70% / 30% Split

The downflow air — which has now swept across the work surface and may carry biological agents from the work in progress — mixes with the inflow air at the front intake grille and is drawn into the recirculation and exhaust plenum below the work surface. Here the contaminated air is split: approximately 70% is directed back to the supply HEPA filter for recirculation (the recirculating portion), and approximately 30% is directed to the exhaust HEPA filter for purification before discharge back to the laboratory room (the exhaust portion). This Type A2 split — 70/30 recirculation/exhaust — is the standard for non-volatile chemical applications.

4

Exhaust HEPA Purifies Room Air — Environment Protection

The exhaust portion of the contaminated air passes through the exhaust HEPA filter (H14, 99.997% at 0.3 micron) before being returned to the laboratory room. This exhaust HEPA removes 99.997% of all biological agents — bacteria, viral particles, fungal spores, and aerosolised cells — ensuring that no living biological agent escapes from the BSC interior to the surrounding laboratory environment. The laboratory personnel working outside the BSC are protected from biological agents generated inside it throughout the working day.

5

Microprocessor Monitors Safety — Alarm Before Operator Awareness

The microprocessor controller continuously samples both inflow and downflow air velocity sensors at a high frequency — detecting a drop in inflow velocity (which would compromise personnel protection) before the velocity has dropped far enough for the operator to notice a change in the airflow sensation at the sash opening. When inflow velocity falls below the safe threshold (0.38 m/s), an audible alarm and visual alert activate immediately — alerting the operator to stop work, close the sash, and investigate the cause before any biological material has escaped the cabinet.

Classification Guide

BSC Classes & Types — Choosing the Right Level of Containment

The type of BSC required depends on the biosafety level of the organisms being handled, whether chemical vapours are present, and whether product protection is needed:

Class I

Personnel & Environment — No Product Protection

Inward airflow only. Protects the operator and the room from bio-aerosols. No HEPA supply — unfiltered room air flows over the work surface. Suitable for general microbiology where sample sterility is not required. Rarely used in modern pharmaceutical laboratories.

Class II Type A2 — Most Common — Metrolabs Standard

Personnel + Product + Environment — BSL-1 and BSL-2

Recirculates ~70% of air through supply HEPA, exhausts ~30% through exhaust HEPA. Inflow ≥0.38 m/s. Grade A work surface. Used in ~90% of all BSC installations worldwide. Suitable for cell culture, microbiology, pharmaceutical QC, hospital labs. Standard for BSL-1 and BSL-2. EN 12469 / NSF 49 certified.

Class II Type B2 — Volatile Chemicals

100% Total Exhaust — Chemical + Biological Dual Hazard

100% of cabinet air is exhausted through a dedicated external duct — no recirculation. Required when volatile toxic chemicals, radionuclides, or carcinogens are used alongside biological work inside the cabinet. Requires external ducting installation — not a plug-and-play cabinet.

Class III

Gas-Tight Glovebox — BSL-3/4 Maximum Containment

Completely sealed, gas-tight enclosure under negative pressure. Operator works through heavy-duty rubber gloves attached to ports. All supply air through two HEPA filters in series. Exhaust through two HEPA filters in series. For BSL-3 and BSL-4 — highest-risk pathogens (Ebola, SARS, anthrax).

Metrolabs Biosafety Cabinet Class II – ergonomic sloped sash, large work zone, microprocessor display, LED illumination
Design Detail

Ergonomic Design — Built for Extended Biological Work Sessions

Biosafety cabinets are used for hours on end. Metrolabs BSC Class II units incorporate ergonomic design features that reduce operator fatigue, improve visibility, and maintain the protective air curtain throughout extended work sessions:

⬇️
10° Sloped Front SashEnables a better seated position and wider field of view into the work zone without requiring the operator to lean forward — maintaining distance from the protective air curtain.
🔧
Recessed Air Intake GrilleThe front air intake grille is recessed below the sash opening — preventing the operator’s forearms from blocking the inward air curtain when working at the front of the work surface.
💡
LED Task LightingBright, low-heat LED lighting illuminates the entire work surface uniformly — no shadows, no heat emission that affects temperature-sensitive cultures, and no UV component that would compromise personnel.
💻
Forearm Support / ArmrestAn ergonomic forearm support above the intake grille allows the operator to rest their forearms during extended pipetting tasks without fatigue and without placing weight on the grille that would obstruct airflow.
Standard Features

Six Engineering Features of the Metrolabs BSC Class II

Every Metrolabs Biosafety Cabinet Class II is built to these standards — making it a precision primary containment instrument, not simply a ventilated enclosure:

🔌

Dual H14 HEPA Filtration — 99.997% — Supply + Exhaust

Two H14 HEPA filters (EN 1822 rated — 99.997% minimum efficiency at 0.3 micron MPPS) are installed in the BSC: one on the supply air path (delivers clean air to the work zone) and one on the exhaust air path (purifies air before returning it to the laboratory). Both are installed in gel-seal or knife-edge frames that prevent any unfiltered air from bypassing the filter perimeter. A factory DOP/PAO integrity test is performed on both HEPA filters before the unit is shipped — sampling multiple points across each filter face — and the test report is supplied as part of the IQ documentation. Differential pressure gauges across both filters provide visible filter loading indicators. A G4 pre-filter in the intake path extends HEPA service life by capturing larger particles before they reach the H14 membrane.

H14 SupplyH14 Exhaust99.997%Gel-Seal FrameDOP Factory Test
🔧

Coved SS Interior — Zero-Joint, Single-Piece Construction

The internal work zone of the Metrolabs BSC is fabricated from surgical-grade SS 304 (standard) or SS 316L (optional) with a single-piece coved interior — all internal corners and floor-to-wall junctions radiused to a continuous large radius, with no internal weld joints or seam lines. The standard point of microbial colonisation in a BSC interior is the weld bead at the floor-to-wall junction — the rough weld bead surface provides a physical anchoring site that is partially protected from the wipe-down disinfection cloth. The single-piece coved pressed construction eliminates this joint and its weld bead entirely, leaving a continuous smooth surface at every corner. The electropolished or mirror-polished internal surface (Ra <0.5μm) makes contamination (media residue, disinfectant film, spill traces) immediately visible under the LED work zone lighting — supporting visual inspection-based cleaning verification.

SS 304 / 316LSingle-Piece CovedEP Ra <0.5μmVHP Compatible
💻

Microprocessor Controller — Real-Time Safety Monitoring with Velocity Alarm

The microprocessor-based controller samples inflow and downflow air velocity sensors continuously — at a frequency that detects a transient airflow anomaly within seconds. The inflow velocity alarm activates when inflow falls below the NSF 49 minimum threshold (0.38 m/s / 75 fpm), producing an audible alert and a visual warning on the front display before the operator is aware of any change in the air sensation at the sash. This is the most critical safety feature — it catches a blower failure or extreme filter loading event before any biological material escapes the cabinet. The display provides real-time readings of inflow velocity, downflow velocity, HEPA filter differential pressure (loading indicator), UV-C lamp cumulative hours, and system status. A USB data port allows logging of operational parameters for GMP environmental monitoring and audit records.

Velocity AlarmΔP DisplayUV HoursUSB Log Port

UV-C Germicidal Lamp — Safety-Interlocked Automated Decontamination

The integrated UV-C germicidal lamp is mounted to provide full coverage of the BSC work surface and the side walls of the work zone during idle periods between work sessions. The UV-C lamp cannot be activated when the sash is in the open operating position — the sash-position interlock ensures that personnel are never exposed to UV-C radiation from the lamp. If the sash is opened during an active UV-C cycle, the lamp deactivates immediately. The microprocessor controller tracks cumulative UV-C lamp hours and alerts when the lamp approaches end-of-life (typically 7,000–9,000 hours), allowing planned replacement before germicidal output falls below the effective threshold. The UV-C lamp provides surface decontamination of the areas it can reach during idle periods — it supplements, but does not replace, the manual IPA wipe-down that is required at the start and end of every work session.

Sash-InterlockedAuto-DeactivateHour CounterProgrammed Timer
🔥

VHP Compatibility — All Components Tested for Room-Wide Decontamination

All internal components of the Metrolabs BSC — including the silicone sash sealing strips, the front viewing panel glazing compound, the UV-C lamp end caps, the HEPA filter housing gaskets, and the work surface drain valve seals — are specified and tested to withstand room-wide Vaporised Hydrogen Peroxide sterilisation cycles at 1000 ppm for standard pharmaceutical VHP cycle durations. For microbiology laboratories that use room-wide VHP decontamination as part of their validated periodic decontamination programme, this compatibility ensures the BSC can be processed in the same VHP cycle as the room without equipment damage — and the laboratory can be returned to service immediately after aeration without any BSC-specific post-VHP inspection or component replacement activity.

1000 ppm VHPAll GasketsViewing PanelDrain Seals
🔋

Vibration-Isolated Blower — Stable Work Surface for Precision Biology

The motor-blower is mounted with vibration-isolation suspension — calibrated coil springs and damping washers that decouple the blower vibration from the cabinet body and the work surface. This vibration isolation is essential for two classes of biological work frequently performed inside a BSC: analytical balance weighing of culture media components (where work surface vibration above 0.01g affects measurement repeatability) and high-magnification optical microscopy of slides and cultures (where vibration above approximately 0.5 micron amplitude causes image blur at magnifications above 200x that makes cell identification unreliable). The vibration-isolated blower also reduces the acoustic noise level at the operator position, contributing to the quiet laboratory environment required for extended concentrated work.

Spring IsolationBalance StableLow AcousticMicroscopy OK
Technical Specifications

BSC Class II — Standard Technical Specification

Metrolabs BSC Class II standard parameters — customisable at every dimension and feature:

SpecificationDetails
ClassificationClass II Type A2
Compliance Standard✓ CertifiedEN 12469 / NSF/ANSI 49
Inflow Velocity≥ 0.38 m/s (75 fpm) — NSF 49 minimum
Downflow Velocity0.25 – 0.38 m/s — Grade A ISO Class 5
Air Recirculation Split~70% recirculated / ~30% exhausted
Supply HEPAH14 — 99.997% at 0.3μm — EN 1822
Exhaust HEPAH14 — 99.997% at 0.3μm — EN 1822
Interior MOCSS 304 std / SS 316L optional — coved
Front Sash10° sloped safety glass — recessed grille
UV-C LampStandard — sash-interlocked — programmable timer
ControllerMicroprocessor — velocity alarm — USB log
Standard Widths900 / 1200 / 1500 / 1800 mm
Power Supply220 V / 50 Hz
Full Specification
ClassClass II Type A2
StandardEN 12469 / NSF/ANSI 49
Inflow Min≥0.38 m/s (75 fpm)
Downflow0.25–0.38 m/s Grade A
Recirculation~70% supply / ~30% exhaust
HEPA FiltersH14 — 99.997% — supply + exhaust
G4 Pre-FilterIntake path — extends HEPA life
Interior MOCSS 304 std / SS 316L opt
Interior FinishEP Ra <0.5μm / Mirror
Sash10° sloped — recessed grille
UV-CStandard — sash-interlocked
ControllerMicroprocessor + USB log
Standards & Compliance
EN 12469NSF/ANSI 49Class II A2EN 1822 HEPAWHO LBMASTM A967GMP / ISO
Optional Configurations

Optional Upgrades & Configurations

🧱

SS 316L Interior Upgrade

Upgrade from SS 304 to SS 316L for superior corrosion resistance against VHP and chloride disinfectants in Grade A/B pharmaceutical BSC applications.

📈

Height-Adjustable Stand

Motorised or manual height-adjustable stand for sitting or standing operation — with integrated ergonomic footrest — for operator comfort during extended biological work sessions.

🔍

PAO / DOP Test Ports

Dedicated upstream and downstream sampling ports for PAO filter integrity test during commissioning and routine ISO/GMP annual certification — required for qualification documentation.

💾

USB / BMS Integration

USB port for operational parameter data download — or Ethernet/BMS integration to relay airflow velocity, filter pressure, and UV hours to the facility building management or environmental monitoring system.

🕐

UV-C Hour Meter

Cumulative UV-C lamp hour counter on microprocessor display for planned replacement scheduling — ensuring germicidal output remains above effective threshold throughout lamp service life.

Exhaust Canopy Connection

Thimble connection canopy allowing the BSC to be connected to the building exhaust system for work involving small quantities of volatile chemicals alongside BSL-1/2 biological agents (Type A2 canopy configuration).

📋

Calibration Certificate

Traceable calibration certificate for the airflow velocity sensor and display — required for GMP instrument qualification documentation at installation and annual recertification.

🚨

Sash Open Alarm

Audible alarm activated when sash is raised above the safe working height — alerting operator that the protective air curtain may be compromised before biological work commences.

Applications

BSC Class II Applications Across Every Regulated Biological Laboratory

EN 12469 / NSF 49 Class II A2 Biosafety Cabinets for every environment where three-way protection — personnel, sample, and environment — is required simultaneously:

💊

Pharmaceutical QC Microbiology

BSC Class II for sterility testing, microbial limit testing, environmental monitoring plate incubation, and cell culture work in pharmaceutical QC microbiology laboratories. GMP IQ documentation at handover.

🧪

Cell Culture & Biotech

BSC Class II for mammalian cell culture, hybridoma work, viral vector production, and primary cell isolation — where simultaneous protection of the operator from the biological agent and protection of the culture from contamination is the primary engineering requirement.

BSL-2 Microbiology

BSC Class II for work with moderate-hazard agents — Staphylococcus, Salmonella, hepatitis B, HIV — at Biosafety Level 2. Inward air curtain protects personnel from aerosol. Exhaust HEPA protects laboratory environment. WHO LBM and CDC/NIH BMBL compliant.

🏥

Hospital & Clinical Microbiology

BSC Class II for clinical microbiology sample processing — blood cultures, sputum specimens, MRSA/VRE cultures, and infectious sample handling. Personnel protection from patient-derived biological agents. NABH compliant throughout.

🍽️

Food Microbiology Testing

BSC Class II for FSSAI food microbiology — pathogen testing (Salmonella, Listeria, E. coli O157), sterility confirmation, and microbiological quality testing of food products. Three-way protection for FSSAI-accredited testing laboratories.

🔬

DNA & PCR Work

BSC Class II for DNA recombinant work, gene therapy research, and PCR preparation requiring both the operator and the sample to be protected from cross-contamination. The downflow clean air prevents amplified DNA from contaminating subsequent PCR reactions.

💊

Pharmacy Drug Compounding

BSC Class II for aseptic pharmaceutical compounding — providing the sterile Grade A environment required for non-hazardous drug compounding while protecting compounding personnel from biological contamination of the product being prepared.

🔬

Research & NABL Labs

BSC Class II for NABL-accredited biotechnology and microbiology research laboratories. Annual NSF-certified airflow recertification. VHP compatible. Full NABL qualification documentation package including DOP test records and calibration certificates.

The Metrolabs Advantage

High-Accuracy Fabrication for Long-Term Biosafety Performance

Metrolabs focuses on the high-accuracy fabrication, advanced control systems, and comprehensive documentation that ensures your BSC meets global biosafety standards throughout its service life:

🔌

Factory DOP Test on Both HEPA Filters Before Dispatch

Every Metrolabs BSC undergoes a factory DOP/PAO integrity test on both the supply HEPA filter and the exhaust HEPA filter — confirming zero bypass at the filter perimeter before the unit leaves the factory. The test report is included in the IQ documentation package. A second DOP test at site installation confirms the HEPA integrity has been maintained through shipping and installation.

🌬

Certified Airflow Velocity — Inflow + Downflow Confirmed at Handover

Airflow velocity measurements at multiple work zone positions (inflow at the sash plane, downflow at the work surface at multiple grid points) are performed and documented at handover — confirming that both the personnel protection threshold (≥0.38 m/s inflow) and the Grade A downflow velocity are achieved and distributed uniformly across the work zone before the unit enters service.

🔥

VHP Compatible Throughout — Every Component, Every Cycle

All components — gaskets, sash seal, UV-C end caps, HEPA housing seals, drain valve — are tested for VHP compatibility at 1000 ppm. Microbiology laboratories using VHP room decontamination can process the BSC in the same cycle as the room, returning to service after aeration without any BSC disassembly.

📝

Annual Recertification Support — NSF Annual Inspection Requirement

BSCs must be annually recertified by a trained third-party NSF/EN certifier — the inspection confirms that airflow velocities remain within specification and that the HEPA filters remain intact. Metrolabs provides the DOP test ports (optional) and the calibration certificates required by the recertification procedure, and can coordinate with accredited recertification services for annual inspection scheduling.

💡 Technical Insight — Annual Airflow Visualisation (Smoke Test)
To maintain the highest level of biosafety, Metrolabs recommends performing an Airflow Visualisation (Smoke Test) as part of the annual NSF recertification. A smoke tracer introduced at the sash opening confirms that the inward air curtain maintains a clean sweep across the full width of the sash without any turbulent eddy that could carry bio-aerosol outward. Smoke at the perimeter of the downflow zone reveals any disruption to the laminar pattern caused by items on the work surface that obstruct airflow — a common finding when tall flasks or stacked equipment create turbulent wakes that could allow cross-contamination between adjacent cultures. The annual smoke test catches the airflow patterns that the velocity sensors alone cannot confirm — because velocity is measured at fixed points, not across the full three-dimensional work zone geometry.

BSC Class II vs Laminar Air Flow (LAF) — Key Differences

Feature 🧪 BSC Class II 🔌 LAF Unit
Protects operator
✓ Inward air curtain
✕ No protection
Protects sample
✓ HEPA downflow
✓ HEPA unidirectional
Protects environment
✓ HEPA exhaust
✕ Air flows to room
Infectious agents
✓ BSL-1 and BSL-2
✕ Never — hazardous
Compliance standard
EN 12469 / NSF 49
ISO 14644-1 Class 5
Typical applications
Pathogens, cell cultures, BSL work
Media prep, PCR, electronics

A Laminar Air Flow unit must never be used for work with infectious agents, pathogens, or biological material with aerosol hazard potential — it does not protect the operator. Use a BSC Class II whenever the material being handled has any potential to cause infection or allergic sensitisation through the respiratory route. See LAF Units →

Supply & Commissioning

From Specification to Factory-Tested & IQ-Documented BSC

Metrolabs manages the complete BSC supply and commissioning process:

01
📋

Specification & Biosafety Assessment

Biosafety level confirmed (BSL-1 / BSL-2). Agents handled reviewed. MOC (SS 304 / SS 316L), cabinet width (900–1800mm), optional features (exhaust canopy / PAO ports / height-adjustable stand / BMS), and room ventilation requirements confirmed. EN 12469 / NSF 49 compliance requirements confirmed. Dimensioned drawing produced and signed off.

02
🔌

Precision Fabrication

Cabinet body fabricated — SS 304 or SS 316L coved single-piece interior. HEPA plenum and blower assembly installed with vibration mounts. Supply and exhaust H14 HEPA filters installed in gel-seal frames. G4 pre-filter installed. Microprocessor controller wired. UV-C lamp mounted with sash interlock. LED task lighting fitted. Sloped sash with recessed grille assembled. Passivated + polished.

03
🌬

Factory Test & Validation

Airflow velocity measured — inflow confirmed ≥0.38 m/s · downflow confirmed 0.25–0.38 m/s · DOP/PAO integrity test on both supply and exhaust HEPA filters · UV-C sash interlock tested · Microprocessor alarm tested at threshold · USB data log verified · Smoke visualisation test at factory · All results documented and test report issued with unit.

04

IQ Documentation Package

SS MTCs, ASTM A967 passivation cert, H14 EN 1822 filter cert, DOP/PAO test report (supply + exhaust HEPA), inflow and downflow velocity measurement grid, microprocessor alarm test record, UV-C interlock test record, smoke test record, calibration certificates (if applicable), and installation photographs compiled for GMP/ISO laboratory IQ documentation at handover.

Visit Us
Metrolabs — Chennai, Tamil Nadu
📍

Factory & Office

4, SIDCO-MICRO, Thirumudivakkam,
Chennai – 600 132, Tamil Nadu
🗺 Open in Google Maps ↗

🌎

Website & GSTN

www.metrolabs.biz
www.schoollabfurniture.com
GSTN: 33AASFM3382K1ZT

Get a Free Quote

Ready to Install a Biosafety Cabinet Class II in Your Laboratory?

Contact Metrolabs for a free biosafety assessment. Our specialists will confirm your BSL level, agent type, cabinet width, optional features, and documentation requirements — then fabricate, factory-validate, install, and commission your BSC with complete IQ documentation at delivery.

🔌
Dual H14 HEPA — 99.997% Supply + Exhaust — Factory DOP Test
Grade A downflow · HEPA exhaust · EN 1822 certification
🌬
EN 12469 / NSF 49 — Inflow ≥0.38 m/s — Velocity Alarm
Microprocessor continuous monitoring · Alarm before awareness
🔧
Coved SS 304/316L Interior — Zero-Joint, Zero-Ledge
Single-piece pressed · EP Ra <0.5μm · VHP 1000 ppm
📋
Audit-Ready — DOP Test + Velocity Grid + IQ Documentation
Both HEPA filters tested · GMP/ISO qualification ready

🧪 Request a Free BSC Class II Quote

Quote Request Received!

Our BSC team will contact you within 24 working hours.

Immediate: 9840931231